5 CALIBRATION AND CALIBRATION KIT
123
5.2.6 Full Two-Port Calibration
Full two-port calibration combines two one-port calibrations for each test port with
measurement of transmission and reflection of a THRU standard in both directions (See
figure 5.11). You can also perform isolation calibration by measuring two LOAD
standards.
Figure 5.11 Full two-port calibration
Before starting calibration perform the following settings: select active channel, set the
parameters of the channel (frequency range, IF bandwidth, etc), and select the
calibration kit.
To open full two-port calibration submenu, use the
following softkeys:
Calibration > Calibrate > Full 2-Port Cal
Port 2
SHORT
LOAD
OPEN
Port
1
Port
2
(optional)
Port
1
Port
2
LOAD
Port 1
SHORT
OPEN