5 CALIBRATION AND CALIBRATION KIT
132
Table 5.6 Trace error correction status
Symbols Definition
RO OPEN response calibration
RS SHORT response calibration
RT THRU response calibration
OP One-path 2-port calibration
F1 Full 1-port calibration
F2 Full 2-port or TRL calibration
5.2.12 System Impedance Z0
Z0is the system impedance of a measurement path. Normally it is equal to the
impedance of the calibration standards, which are used for calibration. The Z0value
should be specified before calibration, as it is used for calibration coefficient
calculations.
To set the system impedance Z0,use the
following softkeys:
Calibration > System Z0
Note Selection of calibration kit automatically determines the
system impedance in accordance with the value specified
for the kit.