5 CALIBRATION AND CALIBRATION KIT
136
To open the table of calibration kits, use the following
softkeys:
Calibration > Cal Kit
To edit a calibration kit, highlight its line in the table.
The calibration kit editing is comprised of two main procedures: defining of calibration
standard (sections 5.3.2 and 5.3.3) and assignment of classes to calibration standards
(section 5.3.4). First you need to perform defining of the calibration standards, and then
the assignment of classes to them. Defining of calibration standards and assignment of
classes to them is performed in different tables.
The label of a calibration kit and its description can be edited in the table of the
calibration kits (figure 5.15). The label appears on the calibration menu softkeys. The
description is just for information of the user.
The table also contains display-only fields: flags of predefined and modified calibration
kits and the counter of the calibration standards in a kit.
5.3.1.1 Calibration Kit Selection for Editing
Move the highlighting to the required line in the calibration kit (See figure 5.15) table
using “” and “ “ arrows.
Note The checkmark in the “Select” field does not matter for
the kit selection for editing, it selects the calibration kit
for calibration.
5.3.1.2 Calibration Kit Label and Description Editing
Move the highlighting to the required line in the calibration kit (See figure 5.15) table
using “” and “” arrows and click on the «Enter» softkey. Then enter the new text in
the table.
To activate the on-screen keyboard, click On-Screen
Keyboard softkey.