5 CALIBRATION AND CALIBRATION KIT
139
5.3.2.1 Standard Adding to Calibration Kit
To add a calibration standard to the table of calibration
standard definition (See figure 5.16), use the following
softkey:
Add STD
5.3.2.2 Standard Deleting from Calibration Kit
To delete a calibration standard from the table of
calibration standard definition (See figure 5.16), use the
following softkey:
Delete STD
5.3.2.3 Calibration Standard Editing
Moving in the table of calibration standard definitions (See figure 5.16) using
navigation keys, enter the parameter values for a calibration kit:
Standard No. The calibration standard number is specified in the
calibration kit data sheet (just for information).
Standard Type Select the standard type:
§Open,
§Short,
§Load,
§Thru/Line,
§Unknown Thru,
§Sliding Load,
§Data-Based
Standard Label Standard labels specified on the calibration menu
softkeys.
Freq. Min. Lower cutoff frequency of the standard.
Freq. Max. Upper cutoff frequency of the standard.
Offset Delay Offset delay value in one direction (s).