5 CALIBRATION AND CALIBRATION KIT
140
Offset Z0 Offset wave impedance value (Ω).
Offset Loss Offset loss value (Ω/s).
Load Impedance Lumped load impedance value (Ω).
C0 1015 F For an OPEN standard, C0coefficient in the polynomial
formula of the fringe capacitance:
C = C0 + C1f + C2f2 + C3f3
C1 1027 F/Hz For an OPEN standard, C1coefficient in the polynomial
formula of the fringe capacitance.
C2 1036 F/Hz2For an OPEN standard, C2coefficient in the polynomial
formula of the fringe capacitance.
C2 1045 F/Hz3For an OPEN standard, C3coefficient in the polynomial
formula of the fringe capacitance.
L0 1012 H For a SHORT standard, L0coefficient in the polynomial
formula of the residual inductance:
L = L0 + L1f + L2f2 + L3f3
L1 10–24 H/Hz For a SHORT standard, L1coefficient in the polynomial
formula of the residual inductance.
L2 10–33 H/Hz2For a SHORT standard, L2coefficient in the polynomial
formula of the residual inductance.
L2 10–42 H/Hz3For a SHORT standard, L3coefficient in the polynomial
formula of the residual inductance.
5.3.2.4 Calibration Standard Copy/Paste Function
To save a calibration standard into clipboard, highlight the
required line in the calibration standard definition table
(See figure 5.16) and click the following softkey:
Copy STD
or
Copy All STDs