5 CALIBRATION AND CALIBRATION KIT
146
5.3.4.2 Standard Deleting from Standard Class Table
Moving in the table of calibration standard classes (See figure 5.18) using navigation
keys, click «Enter» in the required cell for the pop-up menu. Select the line None in the
pop-up menu to delete the standard contained in the cell.
To delete all the standards in the table of calibration
standard classes, use the following softkey:
Clear All CLSs
5.3.4.3 Strict Class Assigment Function
This function allows for limitation of the one standard type(s) available in each class by
the feature of strict correspondence (See table 5.7). If this function is disabled, any class
can be assigned to the standard.
Table 5.7 Standard class and standard type correspondence
NStandard Class Standard Type
1OPEN ·Open,
·Data-Based (One Port)
2SHORT ·Short,
·Data-Based (One Port)
3LOAD
·Load,
·Sliding Load,
·Data-Based (One Port)
4THRU ·Thru/Line,
·Data-Based (Two Port)
5TRL THRU ·Thru/Line,
·Data-Based (Two Port)
6TRL REFLECT
·Open,
·Short,
·Data-Based (One Port)
7TRL LINE/MATCH ·Load,
·Thru/Line