5 CALIBRATION AND CALIBRATION KIT
149
5.4.2 Power Calibration Procedure
Perform connection and setting of an external power meter as described in section 8.10.
Connect the sensor to one of the test ports of the Analyzer and perform calibration as
described below. Then repeat the calibration for the other test port.
To select the calibrated port number, use the following
softkeys:
Calibration > Power Calibration > Select Port
To zero power meter, use the following softkeys:
Calibration > Power Calibration > Power Sensor Zero
Correction
Note Power meter sensor can be connected to the port, as
during zero setting the output signal of the port is turned
off.
To execute power calibration, use the following softkeys:
Calibration > Power Calibration > Take Cal Sweep
Note After the power calibration is complete, power correction
automatically turns on.
5.4.3 Power Correction Setting
To enable/disable power correction, use the
following softkeys:
Calibration > Power Calibration > Correction
5.4.4 Loss Compensation Table Editing
If you need to apply the loss compensation, enable this function and fill out the table
before you start the power calibration procedure. Fill out the table for each port
individually.