5 CALIBRATION AND CALIBRATION KIT
153
5.6 Scalar Mixer Calibration
Scalar mixer calibration is the most accurate method of calibration applied to
measurements of mixers in frequency offset mode.
The scalar mixer calibration requires OPEN, SHORT, and LOAD standards as well as
external power meter (See figure 5.20). The power meter connection and setup is
described in section 8.8.
Figure 5.20 Scalar mixer calibration setup
Port 2
Power
Meter
Power
Meter
Port 1
Port 2
SHORT
LOAD
OPEN
Port
1
Po
rt
2
LOAD
Port 1
OPEN