5 CALIBRATION AND CALIBRATION KIT
155
Click Reflection Port 2 softkey.
Connect SHORT, OPEN and LOAD standards to port 2 as
shown in figure 5.20. Perform two measurements over two
frequency ranges (Freq 1 and Freq 2) for each standard
using the respective standard softkeys.
If the frequency offset is disabled, after measurement over
one frequency range the result will be automatically saved
for the both frequency ranges.
The instrument status bar will indicate Calibration in
progress... when the measurement is in progress. On
completion of the measurement, a check mark will appear
in the left part of the softkey.
Click Transmission softkey.
Connect a THRU standards between the test ports. If the
port connectors allow, connect the ports directly together
(through line with zero electrical length). Perform two
measurements over two frequency ranges (Freq 1 and Freq
2).
If the frequency offset is disabled, after measurement over
one frequency range the result will be automatically saved
for the both frequency ranges.
The instrument status bar will indicate Calibration in
progress... when the measurement is in progress. On
completion of the measurement, a check mark will appear
in the left part of the softkey.
Click Power softkey.
To zero power meter, click Power Sensor Zero Correction
softkey.
Note Power meter sensor can be connected to the port, as
during zero setting the output signal of the port is turned
off.