5 CALIBRATION AND CALIBRATION KIT
156
Connect the power meter to port 1. Perform two
measurements over two frequency ranges (Freq 1 and Freq
2).
If the frequency offset is disabled, after measurement over
one frequency range the result will be automatically saved
for the both frequency ranges.
Connect the power meter to port 2. Perform two
measurements over two frequency ranges (Freq 1 and Freq
2).
The instrument status bar will indicate Calibration in
progress... when the measurement is in progress. On
completion of the measurement, a check mark will appear
in the left part of the softkey.
To complete the calibration procedure, click Apply.
This will activate the process of calibration coefficient
table calculation and saving it into the memory. The error
correction function will also be automatically enabled.
To clear the measurement results of the standards, click
Cancel.
This softkey does not cancel the current calibration. To
disable the current calibration turn off the error correction
function.
Note You can check the calibration status in channel status bar
(See table 5.5) or in trace status field (See table 5.6) –
SMC label.