5 CALIBRATION AND CALIBRATION KIT
165
Select manual or automatic orientation of the ACM
using Orientation softkey.
It is recommended to select AUTO orientation.
Perform confidence check using Confidence Check
softkey.
After the measurement is completed, two traces for each S-parameter will be displayed.
The measured parameters will be shown as the data trace, and the ACM parameters will
be shown as the memory trace.
Compare the data trace and the memory trace of the same parameter, for example S21.
To perform more accurate comparison, enable the function of math operations between
data and memory traces. In the logarithmic magnitude or phase format use the Data /
Memory operation. In the linear magnitude format use the Data – Memory operation.
The conclusion on whether the current calibration provides sufficient accuracy or not is
made by the user.