6 MEASUREMENT DATA ANALYSIS
203
Table 6.7 Time domain gating window shapes
Window Shape Bandpass
Sidelobe Level
Gate Resolution (Minimum Gate
Span)
Minimum – 48 dB
min
max
2.8
F
F
-
Normal – 68 dB
min
max
5.6
F
F
-
Wide – 57 dB
min
max
8.8
F
F
-
Maximum – 70 dB
min
max
25.4
F
F
-
6.5.1 Time Domain Gate Activating
To enable/disable the time domain gating function: toggle
the following softkey:
Analysis > Gating > Gating
Note Time domain gating function is accessible only in linear
frequency sweep mode.
6.5.2 Time Domain Gate Span
To define the span of time domain gate, you can set its start and stop, or center and span
values.
To the start and stop of the time domain gate, use the
following softkeys:
Analysis > Gating > Start
Analysis > Gating > Stop
To set the center and span of the time domain gate, use
the following softkeys:
Analysis > Gating > Center
Analysis > Gating > Span