8 SYSTEM SETTINGS
233
Table 8.1 Supported power meters
Power Meter Name in
Analyzer
Program
USB
Connection
Type
Additional Software
Rohde&Schwarz
NRP-Z series
Sensors (without
Power Meter)
NRP-Z51
(recommended)
R&S NRP-Z
sensors
R&S NRP-Z4
Adapter
nRohde&Schwarz NRP-Toolkit
nRohde&Schwarz RSNRPZ
Instrument driver
Rohde&Schwarz
NRVS Power
Meter and NRV-
Z51 or NRV-Z4
Sensors
R&S NRVS GPIB/USB
Adapter
nGPIB/USB Adapter driver
nVISA visa32.dll Library
To select the power meter, use the following softkeys:
System > Misc Setup > Power Meter Setup > Power Meter
> R&S NRP-Z sensors (USB) | R&S NRVS (GPIB)
If the power meter has GPIB interface, set the GPIB
board address and the power meter address in the bus,
using the following softkeys:
System > Misc Setup > Power Meter Setup > GPIB Board
System > Misc Setup > Power Meter Setup > GPIB
Address
Sensor Info… softkey checks the connection and settings
of the power meter. It provides sensor type, if the
communication between the Analyzer and the power
meter has been successfully established.