4 MEASUREMENT CONDITIONS SETTING
48
nFixture simulation field displays the integrated status of execution of
this function for S-parameter traces. Fixture simulation includes the
following operations: Z0 conversion, embedding, de-embedding. If
the function is enabled for all the traces, you will see black
characters on gray background. If the function is enabled just for
some of the traces, you will see white characters on red background.
nStimulus start field allows for display and entry of the start
frequency or power, depending on the sweep type. This field can be
switched to indication of stimulus center frequency, in this case the
word Start will change to Center.
nSweep points field allows for display and entry of the number of
sweep points. The number of sweep points can be set from 2 to
10001.
nSweep type field allows for display and selection of the sweep type.
The values of this field are represented in table 4.6.
nIF bandwidth field allows for display and setting of the IF
bandwidth. The values can be set from 1 Hz to 30 kHz.
nPower level field allows for display and entry of the port output
power. In power sweep mode the field switches to indication of CW
frequency of the source.
nAveraging status field displays the averaging status if this function is
enabled. The first number is the averaging current counter value, the
second one is the averaging factor.
nStimulus stop field allows for display and entry of the stop frequency
or power, depending on the sweep type. This field can be switched
to indication of stimulus span, in this case the word Stop will change
to Span.