4 MEASUREMENT CONDITIONS SETTING
67
To set the frequency axis display mode, click Segment
Display softkey and select Freq Order or Base Order
option.
4.5.9 Measurement Delay
Measurement delay function allows to add additional time interval at each measurement
point between the moment when the source output frequency becomes stable and the
moment of measurement start. This capability can be useful for measurements in
narrowband circuits with transient period longer than the measurement in one point.
To set the time of measurement delay, use the following
softkeys:
Stimulus > Meas Delay