4 MEASUREMENT CONDITIONS SETTING
85
When the averaging function is enabled, the current number of iterations and the
averaging factor, e.g. «9/10», will appear in the channel status bar. The averaging
process is considered stable when the both numbers have become equal.
The averaging should be set for each channel individually. Before you set the
averaging, first activate the channel.
To toggle the averaging function on/off, use the following
softkeys:
Average > Averaging
To set the averaging factor, use the following softkeys:
Averaging > Avg Factor
4.10.3 Smoothing Setting
The smoothing of the sweep results is made by averaging of adjacent points of the trace
determined by the moving aperture. The aperture is set by the user in percent against the
total number of the trace points.
The smoothing does not increase dynamic range of the Analyzer. It does not affect the
average level of the trace, but it reduces the noise bursts.
The smoothing should be set for each trace individually. Before you set the smoothing,
first activate the trace.
To toggle the smoothing function on/off, use the
following softkeys:
Averaging > Smoothing
To set the smoothing aperture, use the following softkeys:
Averaging > Smo Aperture