9
SAFETY INSTRUCTIONS
Carefully read through the following safety instructions before putting the Analyzer into
operation. Observe all the precautions and warnings provided in this Manual for all the
phases of operation, service, and repair of the Analyzer.
The Analyzer must be used only by skilled and specialized staff or thoroughly trained
personnel with the required skills and knowledge of safety precautions.
PLANAR 804/1 complies with INSTALLATION CATEGORY II as well as
POLLUTION DEGREE 2 in IEC61010–1.
PLANAR 804/1 is MEASUREMENT CATEGORY I (CAT I). Do not use for CAT II,
III, or IV.
PLANAR 804/1 is for INDOOR USE only.
PLANAR 804/1 is tested in stand-alone condition or in combination with the
accessories supplied by PLANAR against the requirement of the standards described in
the Declaration of Conformity. If it is used as a system component, compliance of
related regulations and safety requirements are to be confirmed by the builder of the
system.
Never operate the Analyzer in the environment containing inflammable gasses or
fumes.
Operators must not remove the cover or part of the housing. The Analyzer must not be
repaired by the operator. Component replacement or internal adjustment must be
performed by qualified maintenance personnel only.
Never operate the Analyzer if the power cable is damaged.
Never connect the test terminals to mains.
Electrostatic discharge can damage your Analyzer when connected or disconnected
from the DUT. Static charge can build up on your body and damage the sensitive
circuits of internal components of both the Analyzer and the DUT. To avoid damage
from electric discharge, observe the following:
-Always use a desktop anti static mat under the DUT.
-Always wear a grounding wrist strap connected to the desktop anti static mat via
daisy-chained 1 MΩ resistor.
-Connect clamp on the body of the Analyzer to the body of the DUT before
you start operation.
Observe all the general safety precautions related to operation of equipment powered by
mains.
Terms of Use | Privacy Policy | DMCA Policy